Sitemap  |  Home  |  Contact  |  CAS  |      中文
 
 
  ·search
  ·About us
  ·Research
  ·People
  ·Recently Published Papers
  ·News
  ·Papers of Tribology
  ·Papers of Journal of Molecular Catalysis
  ·Papers of Analysis and Testing Technology and Instruments
  ·Resources
  ·Education & Training
  ·Join Us
  ·Societies & Publications
  ·Papers
  ·Living in Gansu
  ·Links
  ·Sitemap
  ·Pictures
  ·Special
  ·LICP in the News
  ·zhangjunyan
  ·yanxingbin
  ·Home
  ·资源库

Location: Home>Papers of Analysis and Testing Technology and Instruments
Exploration the Detection Method Indoor Air of Total Volatile Organic Compounds
2012-03-31 ArticleSource:
Close Text Size: A A A Print

FU Ying1, WANG Yi-jun1, ZHANG Feng-yan2

(1.  College of Chemical and Environment of Qingdao University, Qingdao 266071, China; 2. Qingdao Supervision and Testing Center of Product Quality, Qingdao 266071, China )

Abstract: Total Volatile Organic Compounds(TVOC) blended standard solution was usedto explore the detection method of TVOC by GC and to determine the best chromatographic conditions. The improved process used only 26.1min, which is 25.9 min shorter than that of the original one(52 min) described in GB/T1 8883-2002. The separating degree of each component in the TVOC blended standard solution was greater than 99%. The RSD of precision was between 1.05% and 2.89%. the recovery was between 95% and 102%, andthe detection limit was 0.020 μg.

Key words: gas chromatography; total volatile organic compounds; improvement of chromatographic conditions

E-mail: yijunwangqdu@163.com

Analysis and Testing Technology and Instruments, Vol. 18, No. 1, 2012,47-56

Address: No.18,Tianshui Middle Road,Lanzhou,P.R.China
ZIP Code:730000 Tel: 86-0931-4968009  Fax: 86-0931-8277088
E-mail: webeditor@licp.cas.cn